1. Physical Dimensions & Environmental
Dimensions (W x H x D): 33 x 31.8 x 13.5 cm (13 x 12.5 x 5.3 inches)
Weight: 4.95 kg (10 lbs) for the basic configuration
Operating Temperature: 0°C to 45°C (32°F to 113°F)
Operating Humidity: 15% to 95% relative humidity (non-condensing)
Atmospheric Pressure: 860 to 1060 hPa
Water/Solid Resistance: IPX4 (Water) / IP2X (Solid foreign objects)
2. Display Specifications
Type: High-resolution TFT Color Display
Size: Available in 7-inch or 8-inch diagonal configurations
Resolution: 800 x 600 pixels
Data Displayed: Heart Rate, Lead/Pads indicator, SpO2, AED Prompts, Alarms, Selected/Delivered Energy.
Sweep Speeds: 12.5 mm/sec, 25 mm/sec, and 50 mm/sec
3. Defibrillator (Manual & AED)
Waveform: Truncated Exponential Biphasic technology
Charge Time: Less than 7 seconds to max energy (360J) with a fully charged battery
Manual Energy Settings: 2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, and 360 Joules
Synchronized Mode: Delivers power synchronized to the patient's R-wave (within 60 msec)
Output Accuracy: ±10% or 1J (whichever is greater) at 50 ohms
Electrode Impedance Measurement: 0 to 250 ohms
AED Functionality: * Automatic ECG analysis and charging.
Pre-programmed auto-energy levels (up to 3 shocks).
Voice and visual prompt guidance.
Paddles: Reusable external adult paddles with integrated, slide-off pediatric adapters.
4. ECG Monitoring
Patient Connection: 3-lead cable, 5-lead cable, or paddles
Lead Selections: I, II, III, aVR, aVL, aVF, V, and Paddles
Heart Rate Range: 20 to 300 BPM
Heart Rate Alarms: User-selectable (high/low), visual (flashing at 2Hz), and audible (800 tone).
Frequency Response: * Monitoring mode: 0.3 to 40 Hz
Analysis mode: 0.05 to 150 Hz
AED mode: 0.5 to 30 Hz
Common Mode Rejection Ratio (CMRR): $\ge$ 60dB
5. Power & Battery Setup
Battery Type: High-Capacity Rechargeable Ni-MH (12V)
Battery Capacity: Yields up to 60 full-energy (360J) discharges or 3 to 4 hours of continuous ECG monitoring.
Charge Rate: 80% charge in 7.5 hours; 100% in 9 hours.
AC Power Compatibility: 100V–240V, 50/60 Hz
6. Built-in Thermal Recorder (Printer)
Paper Type/Size: Thermal paper, 50 mm width
Print Head: High-resolution thermal array
Print Speeds: 12.5 mm/s, 25 mm/s, or 50 mm/s
Modes: Manual or Automatic (triggers a 9-second printout upon alarm activation, charge, or shock discharge).
7. Optional Modules (If Upgraded)
SpO2 (Pulse Oximetry): Range of 30% to 100%. Accuracy of ±2% (between 80–90%).
NIBP (Non-Invasive Blood Pressure): Automatic oscillation method. Tracks systolic, diastolic, and mean pressure for adult, pediatric, and neonatal groups.
EtCO2 (End-tidal CO2): Range of 0 to 13% with a measurement accuracy of ±3 mmHg.
Pacemaker: VVI demand pacing, rate variable from 30 ppm to 180 ppm.